Anatomy of the Lens and its Relationship with the Final Position of the Intraocular Lens

Authors

Keywords:

cataract, microsurgery, anatomical planes of the lens

Abstract

Objective: To determine the relationship between the anatomical planes of the lens and the actual position of the intraocular lens in patients diagnosed with cataracts. Methods: A descriptive, longitudinal, retrospective case series study was conducted on 78 eyes of 78 patients who underwent cataract surgery with intraocular lens implantation using phacoemulsification; the SRK/T formula was used. Optical coherence tomography imaging provided by the IOL Master 700 was used to determine the anatomical planes of the lens, as well as the actual position of the intraocular lens in patients treated at the Ocular Microsurgery Center of the Ramón Pando Ferrer Cuban Institute of Ophthalmology during the period January 2018-January 2021.

Results: The group of patients over 60 years of age and male predominated. The estimated spherical equivalent and obtained spherical equivalent were related to the range calculated for both groups according to the refractive error. The prediction error of the formula had a strong correlation between the estimated spherical equivalent and the obtained spherical equivalent. The actual position of the intraocular lens is more related to the equatorial plane of the lens.

Conclusions: The anatomical parameters of the lens are important variables for the development and refinement of formulas for intraocular lens calculation.

Downloads

Download data is not yet available.

Author Biographies

Yaumary Bauza Fortunato, Universidad de Ciencias Médicas de La Habana, Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana, Cuba

Especialista de II grado en Oftalmología

Master en cirugía de segmento anterior y longevidad satisfactoria

Profesora e investigadora Auxiliar

Zucell Ana Veitía Rovirosa, Universidad de Ciencias Médicas de La Habana, Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana, Cuba

Especialista de II grado en Oftalmología

Master en cirugía de segmento anterior y longevidad satisfactoria

Profesora e investigadora auxiliar

Iraisi Francisca Hormigó Puertas, Universidad de Ciencias Médicas de La Habana, Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana, Cuba

Especialista de II grado en Oftalmología

Doctora en Ciencias Médicas

Profesora Titular e Investigadora Auxiliar

María de las Mercedes Dueñas Moreno, Universidad de Ciencias Médicas de La Habana, Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana, Cuba

Licenciada en optometría y óptica

References

1. Tinoco R, Guerrero O, Arroyo L. Posición efectiva del lente en pacientes con síndrome de pseudoexfoliación. Revista Mexicana de Oftalmología. 2011 [acceso 25/10/2023];85(3). Disponible en: https://www.imbiomed.com.mx/articulo.php?id=78420

2. Bauza Y, Plasencia RH, Hernández I, Veitía ZA, Batista AJ. Relación de los planos anatómicos del cristalino con la posición real del lente intraocular. Rev Cubana Oftalmol. 2022 [acceso 25/10/2023];35(4):e1680. Disponible en: http://scielo.sld.cu/scielo.php?script=sci_arttext&pid=S086421762022000400012&Ing=es

3. Flaxman SR, Bourne RRA, Resnikoff S, Ackland P, Braithwaite T, Cicinelli MV, et al. Global causes of blindness and distance vision impairment 1990-2020: a systematic review and meta-analysis. Lancet Glob Health. 2017;5(12):e1221-e34. DOI: 10.1016/S2214-109X

4. Hormigó IF. Invisibilidad de los pacientes diabéticos tipo 2 para el acceso oportuno a la cirugía de catarata. 2021-2022 [tesis]. La Habana; Universidad de Ciencias Médicas; 2023.

5. Boyd K. ¿Qué son las cataratas? USA: American Academy of Ophthalmology; 2023 [acceso 12/01/2024]. Disponible en: https://www.aao.org/salud-ocular/enfermedades/que-son-las-cataratas

6. Bilbao R, Gónzalez F, Llovet A, Ortega J, Fernández VT, Llovet F. Lens-based surgical correction of presbyopia. Where are we in 2020? Arch. Soc. Esp. Oftalmol. 2021;96:(2),74-88. DOI: 10.1016/j.oftal.2020.07.012

7. Yépez C, Zamora M, Reyes V, Rodríguez H, Ramírez K, Ruiz F, et al. Correlación entre la escala LOCS III y la función visual en pacientes con catarata senil. Innovación y desarrollo tecnológico revista digital. 2023 [acceso 12/01/2024];15(1). Disponible en: https://iydt.wordpress.com/wp-content/uploads/2022/12/1_24_correlacion-entre-la-escala-locs-iii-y-la-funcion-visual-en-pacientes-con-catarata-senil.pdf

8. Hu CY, Jian JH, Cheng YP, Hsu HK. Analysis of crystalline lens position. J Cataract Refract Surg. 2006; 32(4): 599-603. DOI: 10.1016/j.jcrs.2006.01.016

9. Meihe L, Ramos Y, Cárdenas T, Dengtan L. Resultado visual y reincorporación laboral de pacientes operados de catarata traumática. Rev Cubana Oftalmol 2024 [acceso 12/01/2024];37(0). Disponible en: https://revoftalmologia.sld.cu/index.php/oftalmologia/article/view/1854

10. Vámosi P, Nemeth G, Berta A. Pseudophakic accommodation with 2 models of foldable intraocular lenses. J Cataract Refract Surg. 2006;32(2):221-6. DOI: 10.1016/j.jcrs.2005.08.059

11. Hernández Y, Hormigó I, Ramos L, Bauza Y, García F. Síndrome de distensión de la bolsa capsular: del diagnóstico al tratamiento. Rev Cubana Oftalmol. 2023 [acceso 12/01/2024];36(1). Disponible en: https://revoftalmologia.sld.cu/index.php/oftalmologia/article/view/1669

12. Dooley I, Charalampidou S, Nolan J, Loughman J, Molloy L, Beatty S. Estimation of effective lens position using a method independent of preoperative keratometry readings. J Cataract Refract Surg. 2011;37(3):506-12. DOI: 10.1016/j.jcrs.2010.09.027

13. Dalmagro JA, Urrets JA. Ecometría para el cálculo de lentes intraoculares. Oftalmol Clin Exp. 2016 [acceso: 15/02/2024];9(3):75-87. Disponible en: https://oftalmologos.org.ar/oce_anteriores/files/original/b6b84f35b1e6af498ab39fd5c8c1f92d.pdf

14. Fernández FL, Segarra J. Cálculo de la lente intraocular: ¿Qué fórmula usar y por qué? Fuentes de error en queratometría y biometría. En: Alió JL, Rodríguez JL. Buscando la excelencia en la cirugía de catarata. España: Editorial Glosa, Barcelona; 2007. p: 66-91.

15. Pérez E, Veitía ZA, Santiesteban I, Montero E, Andújar P, Cuan Y. Cálculo del lente intraocular en la cirugía de catarata. En: Río Torres M, Capote Cabrera A, Padilla González CM, Eguía Martínez F, Hernández Silva JR. Oftalmología. Criterios y tendencias actuales. La Habana: Editorial Ciencias Médicas; 2009. p. 223-43.

16. Tsunehiro S, Shimizu K, Shoji N, Hiro-Oka H, Furukawa H. Prediction of intraocular lens position based on crystalline lens shape measured using anterior segment optical coherence tomography. Kitasato Med J. 2017 [acceso 15/02/2024];47:110-7. Disponible en: https://www.kitasato-u.ac.jp/ktms/kaishi/pdf/KMJ47-2/KMJ47-2p110-117.pdf

17. Martinez-Enriquez E, Sun M, Velasco M, Birkenfeld J, Pérez P, Marcos S. Optical coherence tomography based estimates of crystalline lens volume, equatorial diameter and plane position. Invest Ophthalmol. Vis Sci.2016;57: 600–10. DOI: 10.1167/iovs.15-18933

18. Martinez-Enriquez E, Pérez P, Durán S, Jiménez I, Marcos S. Estimation of intraocular lens position from full crystalline lens geometry: towards a new generation of intraocular lens power calculation formulas. Scientific REPorTS |. 2018;8:9829. DOI: 10.1038/s41598-018-28272-6

19. Plasencia RH. Relación de los planos anatómicos del cristalino con la posición real de la lente intraocular [tesis]. La Habana: Universidad de Ciencias Médicas; 2021.

20. Hormigó Puertas IF, Montero Díaz E, Cuan Aguilar Y. Estudios preoperatorios para el cálculo del poder de la LIO. En: Cárdenas Díaz T. Óptica y optometría. Principios y aplicación clínica. La Habana: Editorial Ciencias Médicas; 2023. [acceso 27/04/2024]. Disponible en: http://www.bvscuba.sld.cu/libro/optica-y-optometria-principios-y-aplicacion-clinica_volumen1

21. Díez Montero C. Relación entre los componentes principales del cristalino y la profundidad de cámara anterior después de la formación de catarata [tesis]. España: Universidad de Valladolid; 2024. Disponible en: https://uvadoc.uva.es/handle/10324/66897

22. World Medical Association. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human participants. JAMA; 2024. DOI: 10.1001/jama.2024.21972

23. Dirección de Registros Médicos y Estadísticas de Salud. Anuario Estadístico de Salud, 2022. MINSAP. 2023 [acceso 05/05/2024]. Disponible en: https://instituciones.sld.cu/ucmvc/files/2023/10/Anuario-Estad%C3%ADstico-de-Salud-2022-Ed-2023.pdf

24. Burton MJ, Ramke J, Marques AP, Bourne RR, Congdon N, Jones I, et al. The Lancet Global Health commission on Global Eye Health: vision beyond 2020. Lancet Glob Health. 2021; 9(4): e489–e551. DOI: 10.1016/S2214-109X(20)30488-5

25. Veitía ZA, Hernández ME, Pérez EC, Rodríguez B, Méndez AM, Hernández I. Resultados del lente intraocular con la fórmula holladay 2 en pacientes con catarata. Rev Cubana Oftalmol 2020 [acceso 05/05/2024];33(3). Disponible en: http://scielo.sld.cu/scielo.php?script=sci_arttext&pid=S0864-21762020000300002

26. Vertía ZA, Simón DC, Pérez EC, Hernández I, Bauza Y. Resultados del cálculo de la lente intraocular con fórmula Holladay 2 y Barret Universal 2. Rev Cubana Oftalmol 2022 [acceso 5/5/2024];35(3). Disponible en: http://scielo.sld.cu/scielo.php?script=sci_arttext&pid=S0864-21762022000300009&lng=es.

27. Mengchan S. Biometric measurements in the crystalline lens: applications in cataract surgery [tesis]. [España]: Universidad de Valladolid. Facultad de Medicina; 2017 [acceso 05/05/2024]. Disponible en: http://uvadoc.uva.es/handle/10324/26558

28. Boletín epidemiológico. 2024 [acceso 05/05/2024];21(4). Disponible en: https://temas.sld.cu/estadisticassalud/2024/03/

29. Jeong J, Song H, Lee JK, Chuck RS, Kwon JW. The effect of ocular biometric factors on the accuracy of various IOL power calculation formulas. BMC Ophthalmol. 2017; 17(1): 62. DOI: 10.1186/s12886-017-0454-y

30. Satou T, Shimizu K, Tsunehiro S, Igarashi A, Kato S, Koshimizu M, et al. Relationship bettween crystalline lens thickness and shape and the identification of anterior ocular segment parameters for predicting the intraocular lens position after cataract surgery. Hindawi BioMed Research International. 2019(9829):3458548. DOI: 10.1155/2019/3458548

31. Yoo YS, Whang WJ, Hwang KY, Lazo M, Hwang JH, Joo CK, Yoon G. Use of the Crystalline Lens Equatorial Plane as a New Parameter for Predicting Postoperative Intraocular Lens Position. Am J Ophthalmol. 2019;198:17-24. DOI: 10.1016/j.ajo.2018.09.005

Published

2025-06-24

How to Cite

1.
Bauza Fortunato Y, Veitía Rovirosa ZA, Garmendía Martínez A, Hormigó Puertas IF, Dueñas Moreno M de las M. Anatomy of the Lens and its Relationship with the Final Position of the Intraocular Lens. Rev Cubana Oftalmol [Internet]. 2025 Jun. 24 [cited 2025 Aug. 28];38. Available from: https://revoftalmologia.sld.cu/index.php/oftalmologia/article/view/2036

Issue

Section

Investigaciones